- Kim Jong Chan
- Department of Electronic Engineering, Yonsei University
-
- Song Chui Gyu
- Biomedical Research Engineering Center, Samsung Medical Center
-
- Lee Jae Chon
- Ajou University and Institute for Advanced Engineering
-
- Oh Myung Hwan
- Korea Institute of Science and Technology
-
- Cha Il Whan
- Department of Electronic Engineering, Yonsei University
この論文をさがす
抄録
It is known that 3-D imaging enhances the ability to perform delicate endoscopic maneuvers, such as dissection or precise suturing. Moreover, it provides observers with a better understanding of spatial anatomy. We propose here a 3-D polarization-type endoscopic imaging system, which has three features of stereo endoscopic image processing features: real time image capture, retrieval, and presentation of left and right images on a TFf-LCD monitor. We also compare the established 2-D type, the electric shutter-type 3-D imaging, and the new polarizationtype 3-D endoscopic imaging systems in terms of the performance of laparoscopic tasks in an animate situation. The polarization-type 3-D system is shown to be superior to the electric shutter-type in image quality, stability of performance, and speed based on the testing of suturing and loop passing.
収録刊行物
-
- bioimages
-
bioimages 6 (3-4), 93-97, 1998
日本バイオイメージング学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390564227304851840
-
- NII論文ID
- 10002038618
-
- NII書誌ID
- AA11084187
-
- ISSN
- 09192719
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可