A Digital Waveform Analysis System with a Fast Digitizer and VXIbus System.
-
- KURAHASHI Tomohiko
- Department of Quantum Engineering and Systems Science Graduate School of Engineering The University of Tokyo
-
- TAKAHASHI Hiroyuki
- Department of Quantum Engineering and Systems Science Graduate School of Engineering The University of Tokyo
-
- NAKAZAWA Masaharu
- Department of Quantum Engineering and Systems Science Graduate School of Engineering The University of Tokyo
Bibliographic Information
- Other Title
-
- Digital Waveform Analysis System with a
Search this article
Abstract
A radiation digital waveform processing system has been studied using a fast signal digitizer on the VXIbus system. The proposed system is able to measure pulse signals at the maximum counting rate of over 160 waveforms per second. When applying to an n-γ pulse shape discrimination for an NE213 scintillation detector, more refined pulse shape discrimination can be easily accomplished by adjusting the discrimination parameter on the basis of energy of each incident radiation.
Journal
-
- Journal of Nuclear Science and Technology
-
Journal of Nuclear Science and Technology 35 (7), 473-476, 1998
Atomic Energy Society of Japan
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679073006592
-
- NII Article ID
- 10002079679
-
- NII Book ID
- AA00703720
-
- COI
- 1:CAS:528:DyaK1cXls1Crsrw%3D
-
- ISSN
- 18811248
- 00223131
-
- NDL BIB ID
- 4525683
-
- Text Lang
- en
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed