無電解プロセスによるグルコースのアンペロメトリックセンサーの作製 Amperometric Glucose Sensor Fabricated Using Electroless Processes

この論文にアクセスする

この論文をさがす

著者

抄録

Nickel, copper and copper oxide films were fabricated by electroless plating and chemical oxidation. The analytical performance of glucose by constant potential amperometry was studied in the flow system with films. An electrocatalytic glucose oxidation peak appeared in cyclic voltammograms of all films, similar to films prepared by other methods. In the flow system, the largest current was observed for the copper oxide film: excellent stability and the widest calibration curve linearity were also seen for the film. Scanning electron microscopy revealed a fine network structure for the copper oxide film. The large surface area due to such structure probably resulted in the large current.

収録刊行物

  • 表面技術 = The Journal of the Surface Finishing Society of Japan  

    表面技術 = The Journal of the Surface Finishing Society of Japan 49(8), 894-899, 1998-08-01 

    The Surface Finishing Society of Japan

参考文献:  20件

参考文献を見るにはログインが必要です。ユーザIDをお持ちでない方は新規登録してください。

各種コード

  • NII論文ID(NAID)
    10002110322
  • NII書誌ID(NCID)
    AN1005202X
  • 本文言語コード
    JPN
  • 資料種別
    ART
  • ISSN
    09151869
  • NDL 記事登録ID
    4546114
  • NDL 雑誌分類
    ZP41(科学技術--金属工学・鉱山工学)
  • NDL 請求記号
    Z17-291
  • データ提供元
    CJP書誌  NDL  J-STAGE 