Chromatographic Simulation of the Effect of a Polyvalent Counter Ion on Retention Behavior and Peak Intensity of Sulfate.
-
- NISHIMURA Masayuki
- Analytical Applications Department, Shimadzu Corporation
-
- HAYASHI Morimasa
- Analytical Applications Department, Shimadzu Corporation
-
- YAMAMOTO Atsushi
- Toyama Institute of Health
-
- HAYAKAWA Kazuichi
- Faculty of Pharmaceutical Sciences, Kanazawa University
-
- MIYAZAKI Motoichi
- Faculty of Pharmaceutical Sciences, Kanazawa University
Search this article
Abstract
The effect of a mobile phase polyvalent counter ion on the retention behavior and peak intensity of sulfate in non- suppressed ion chromatography was quantitatively studied using a chromatogram simulation method. In this study, both indirect photometric detection (IPD) method and conductometric detection (CD) method were evaluated. With the IPD method, the retention of sulfate was observed to decrease in the presence of the polyvalent counter ion (Trien) without any change in peak intensity. With the CD method, however, peak intensity of sulfate decreased and, under some conditions, gave a negative peak in simulation as was also observed in actual experiments. The present work demonstrates the possibility of predicting the effect of a polyvalent counter ion on sulfate retention without resorting to physical experiments.
Journal
-
- Analytical Sciences
-
Analytical Sciences 11 (5), 755-758, 1995
The Japan Society for Analytical Chemistry
- Tweet
Details 詳細情報について
-