回分式静電粒径測定装置の開発 Development of a Batch Type Electrical Mobility Analyzer

この論文にアクセスする

この論文をさがす

著者

抄録

A new batch type electrical mobility analyzer, or electrical sedimentation particle sizer, has been developed and the basic performance has been evaluated. Two types of electrical sedimentation cells, with concentric cylinder electrodes and parallel disk electrodes, are constructed and evaluated. It was found that the size distribution of sodium chloride particles measured by these cells agreed well with that measured by the conventional DMA (Differential Mobility Analyzer). Since this method measures particles in a closed sedimentation cell, it is advantageous for measuring aerosol particles which fluctuate in size and number concentration with time.

収録刊行物

  • エアロゾル研究 = Journal of aerosol research  

    エアロゾル研究 = Journal of aerosol research 10(2), 135-140, 1995-06-20 

    日本エアロゾル学会

参考文献:  7件

参考文献を見るにはログインが必要です。ユーザIDをお持ちでない方は新規登録してください。

各種コード

  • NII論文ID(NAID)
    10002686586
  • NII書誌ID(NCID)
    AN10041511
  • 本文言語コード
    JPN
  • 資料種別
    OTR
  • ISSN
    09122834
  • データ提供元
    CJP書誌  J-STAGE 
ページトップへ