変位計の線形誤差の自律的校正

書誌事項

タイトル別名
  • Self Calibration of the Linearity Error of Displacement Sensors.
  • ヘンイケイ ノ センケイ ゴサ ノ ジリツテキ コウセイ

この論文をさがす

抄録

A new method to calibrate precision meters for metrology is proposed. In the method two meters including sensors and amplifiers are used to compare their outputs with each other and any accurate meter for the reference of calibration is not necessary. The calibration system is constructed by two meters A and B to be calibrated, a kind of lever to give displacement and precision stages positioning the sensors axially. Calibrating with this system, the linearity error of calibration line is determined within an uncertainty limit, which is given by resolution power of the sensors, electric noise and thermal disturbance of circumstance. As a trial experiments, capacitance type displacement meters (micro sense) have been calibrated. The output of the meters with 50 μm range is determined within an uncertainty of about 10 nm. The proposed self-calibration method can determine the most certain calibration curve which can be able to be obtained in the circumstance of the experiment.

収録刊行物

  • 精密工学会誌

    精密工学会誌 59 (12), 2043-2048, 1993

    公益社団法人 精密工学会

被引用文献 (5)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘