Evaluation of film dosimetry for narrow photon beams using linear accelerator

  • Gotoh Shinichi
    Division of Radiological Technology, Nagasaki University Hospital

Bibliographic Information

Other Title
  • 写真フィルムによるナロー一ビームの線量評価

Search this article

Abstract

Narrow photon beam dosimetry used in stereotactic radiosurgery can be made difficult because of the presence of lateral electoronic disequilibrium and steep dose gradiants. Dosimetric characteristics of 10MV circular narrow photon beams ranging in size from 11.6 to 31.4 mm are reported. These characteristics include the measurement of tissue maximum ratio, off axis ratio and relative output factors. Measurements of these parameters were performed in a solid water phantom using x-ray films and cylindrical micro ionization chamber. Although in 50% isodose level width the measurement by micro ionization chamber and x-ray film correlated well, in 80% isodose level width measurement by micro chamber showed apparent underevaluation than that by x-ray film. The relative output factors measurement by micro ionization chamber was consistent with that by x-ray film within 3 % for the field size between 11.6 and 31.4 mm φ.

Journal

Citations (1)*help

See more

References(12)*help

See more

Details 詳細情報について

  • CRID
    1390001204672163968
  • NII Article ID
    10003723851
  • NII Book ID
    AA11137895
  • DOI
    10.11323/jjmp1992.17.1_9
  • ISSN
    21869855
    09188010
  • Text Lang
    ja
  • Data Source
    • JaLC
    • CiNii Articles
  • Abstract License Flag
    Disallowed

Report a problem

Back to top