低エネルギーX線領域におけるセンシトメトリー The X-ray Sensitometry in the Low Energy X-ray Region

この論文にアクセスする

この論文をさがす

著者

    • 天野 貴司 AMANO Takashi
    • 川崎医療短期大学 放射線技術科 Department of Radiological Technology, Kawasaki College of Allied Health Professions
    • 荒尾 信一 ARAO Shinichi
    • 川崎医療短期大学 放射線技術科 Department of Radiological Technology, Kawasaki College of Allied Health Professions
    • 北山 彰 KITAYAMA Akira
    • 川崎医療短期大学 放射線技術科 Department of Radiological Technology, Kawasaki College of Allied Health Professions
    • 山下 一也 YAMASHITA Kazuya
    • 川崎医療短期大学 放射線技術科 Department of Radiological Technology, Kawasaki College of Allied Health Professions

抄録

Inverse square sensitometry in the low energy region is limited by the air attenuation. Therefore, it is difficult to obtain the accurate characteristic curve for the mammography system.<BR>We applied the new method using the technique of Bednarek with the gradual change of the distance to sensitometry in the low energy region. Then we compared it with inverse square sensitometry, which corrected air attenuation by the method already suggested.<BR>As a result, the characteristic curve got by the two different techniques coincided with each other.<BR>It was confirmed that the new method was useful as sensitometry in the low energy region.

収録刊行物

  • 放射線医学物理

    放射線医学物理 18(3), 272-279, 1998-09-30

    Japan Society of Medical Physics

参考文献:  8件中 1-8件 を表示

被引用文献:  3件中 1-3件 を表示