Amperometric Glucose Sensor Based on a Ploydimethyl Siloxane/Enzyme-Bilayer Membrane
-
- MIZUTANI Fumio
- National Institute of Bioscience and Human-Technology
-
- SAWAGUCHI Takahiro
- National Institute of Bioscience and Human-Technology
-
- YABUKI Soichi
- National Institute of Bioscience and Human-Technology
-
- IIJIMA Seiichiro
- National Institute of Bioscience and Human-Technology
書誌事項
- タイトル別名
-
- Amperometric Glucose Sensor Based on a Polydimethyl Siloxane/Enzyme-Bilayer Membrane
この論文をさがす
抄録
<p>An amperometric glucose-sensing electrode based on the cathodic detection of oxygen consumption during the glucose oxidase (GOx) reaction was prepared as follows. A platinum electrode was modified with polydimethyl-siloxane (PDMS), and a GOx-layer was placed on the PDMS layer. The permselectivity of PMDS made it possible to monitor the oxygen consumption rate without interference from the hydrogen peroxide. The concentration of glucose (0.01-1.2 mM) could be determined from the decrease in the cathodic current at −0.1 V vs. Ag/AgCl, without any interference from uric acid, acetaminophen and L-cysteine.</p>
収録刊行物
-
- Electrochemistry
-
Electrochemistry 67 (12), 1138-1140, 1999-12-05
公益社団法人 電気化学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001277392947456
-
- NII論文ID
- 10004275124
-
- NII書誌ID
- AN00151637
-
- ISSN
- 21862451
- 13443542
- http://id.crossref.org/issn/13443542
-
- NDL書誌ID
- 4927361
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ