完全埋込型人工心臓用経皮エネルギー伝送システムの伝導性妨害波の評価・検討 Evaluation of Conducted Emission from Transcutaneous Energy transmission System for a Totally-Implantable Artificial Heart

この論文にアクセスする

この論文をさがす

著者

抄録

This paper reports on evaluation and investigation of conducted emission from transcutaneous energy transmission (TET) system for a totally-implantable artificial heart. Coreless type and externally coupled type of transformers were used for the TET, respectively, and the conducted emissions from the TET system were measured and compared with German regulation of VDE. As a result, the TET system using the externally coupled type of transformer suppressed the conducted emission within the regulation limit in the range from 0.01 to 30MHz.

収録刊行物

  • 電気学会論文誌. D, 産業応用部門誌 = The transactions of the Institute of Electrical Engineers of Japan. D, A publication of Industry Applications Society  

    電気学会論文誌. D, 産業応用部門誌 = The transactions of the Institute of Electrical Engineers of Japan. D, A publication of Industry Applications Society 119(12), 1560-1561, 1999-12-01 