- 高田 義久
- 筑波大学物理工学系
書誌事項
- タイトル別名
-
- Review of beam delivery system for radiotherapy using heavy-charged-particlebeam
この論文をさがす
抄録
Beam delivery techniques for radiotherapy using heavy-charged-particle beam have been reviewed. The conventional techniques using fixed-modulation SOBP (Spread-Out Bragg Peak) are described. The double scattering methods and the wobbling methods are summarized as open systems. The newly-developed conformal techniques,3D spot scanning at PSI,3D raster scanning at GSI, and broad-beam 3D irradiation method at NIRS, are introduced as a method to improve the dose localization by using the variable modulation. It is pointed out that these new systems are the closed system since the dose for each beam configuration and some beam modifying devices should be controlled during the irradiation and that high-level security system is required for highdoserate techniques. Beam delivery techniques for charged particle therapy are rapidly evolving to improve the dose localization further.
収録刊行物
-
- Japanese Journal of Medical Physics
-
Japanese Journal of Medical Physics 19 (1), 16-28, 1999
一般社団法人 日本医学物理学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001204672768896
-
- NII論文ID
- 10004472240
-
- NII書誌ID
- AA11137895
-
- ISSN
- 21869855
- 09188010
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可