Calculated Peak Efficiency of Ge Detector for Volume Source

Bibliographic Information

Other Title
  • 容積線源に対するGe検出器のピーク効率の計算

Search this article

Abstract

A calculation method to determine γ-ray peak efficiency was developed, which was based on numerical integration of peak efficiency calculated by γ-ray path length through a Ge detector. The calculation needs some physical parameters of a Ge detector, such as crystal size and dead layer thickness, sample size and elemental composition of sample matrix, and one or two experimentally obtained peak efficiency for normalization. The method determines peak efficiency including self-absorption correction for a volume source consisted of any kind of matrix.

Journal

  • RADIOISOTOPES

    RADIOISOTOPES 48 (10), 617-625, 1999

    Japan Radioisotope Association

Citations (3)*help

See more

References(29)*help

See more

Details 詳細情報について

Report a problem