蛍光ガラス線量計(SC-1)高エネルギー重イオンに対する応答 Response of Radiophotoluminescent Glass (SC-1) to Hich-energy Heavy lons

この論文にアクセスする

この論文をさがす

著者

抄録

<I>Relative efficiency of a radiophotoluminescent glass (RPLG, product name: SC-1) was examined for</I>He, C, Ne, Si, Ar, <I>and</I>Fe<I>beams and</I> γ-<I>rays from</I><SUP>60</SUP>Co<I>and</I><SUP>137</SUP>Cs<I>sources. Although dose dependency was not seen under 10-100</I>mGy-<I>water exposure, a strong LET dependency was observed. The RPL efficiency decreased almost proportionally with the logarithm of the infinite LET in water from 0.5 to 200</I>keVμm<SUP>-1</SUP>.<I>These results suggest that the value of RPLG obtained as γ-ray equivalent should underestimate the absorbed dose where heavy charged particles are mixed like in space. It is expected, however, to utilize RPLG for the evaluation of radiation quality by combining other detectors having different LET dependencies.</I>

収録刊行物

  • Radioisotopes  

    Radioisotopes 48(10), 645-647, 1999-10-15 

    公益社団法人 日本アイソトープ協会

参考文献:  5件

参考文献を見るにはログインが必要です。ユーザIDをお持ちでない方は新規登録してください。

各種コード

  • NII論文ID(NAID)
    10004565100
  • NII書誌ID(NCID)
    AN00351589
  • 本文言語コード
    ENG
  • 資料種別
    NOT
  • ISSN
    00338303
  • NDL 記事登録ID
    4876538
  • NDL 雑誌分類
    ZM35(科学技術--物理学) // ZN36(科学技術--原子力工学・工業) // ZP2(科学技術--化学・化学工業--物理化学) // ZS45(科学技術--医学--放射線医学)
  • NDL 請求記号
    Z15-20
  • データ提供元
    CJP書誌  NDL  J-STAGE 