Voltammetric Behavior of Adriamycin and Its Determination at Ni Ion-Implanted Electrode.
この論文をさがす
抄録
With a Ni ion-implanted glassy carbon electrode (GCE) as the working electrode in 0.1 mol/l HAc-NaAc (pH=4.62) solution, a sensitive reductive wave of adriamycin was obtained by linear-sweep voltammetry. The peak potential was -0.55 V (vs. SCE). The peak current was proportional to the concentration of adriamycin over the range of 5.2×10-7 - 1.0×10-5 mol/l and the detection limit was 2.7×10-7 mol/l. The behavior of the reduction wave was studied and the wave was applied to the determination of adriamycin in urine. The reduction process was quasi-reversible. AES and XPS experiments showed that Ni was surely implanted into the surface of the GCE, and the implanted Ni at the GCE improved the electrocatalytic activity.
収録刊行物
-
- Analytical Sciences
-
Analytical Sciences 15 (12), 1215-1218, 1999
社団法人 日本分析化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204255795712
-
- NII論文ID
- 130004440021
- 10004708693
-
- NII書誌ID
- AA10500785
-
- COI
- 1:CAS:528:DyaK1MXotFSktbs%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL書誌ID
- 4952753
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可