Imaging of Injection and Separation Processes of DNA on a Microfabricated Capillary Electrophoresis Device
-
- Ueda Masanori
- Department of Medicinal Chemistry, Faculty of Pharmaceutical Sciences, The University of Tokushima Japan Foundation for Aging and Health
-
- Abe Hirohisa
- Technology Research Laboratory, Shimadzu Corporation
-
- Kuyama Hiroki
- Technology Research Laboratory, Shimadzu Corporation
-
- Nakanishi Hiroaki
- Technology Research Laboratory, Shimadzu Corporation
-
- Arai Akihiro
- HPLC Business Department, Analytical Instruments Division, Shimadzu Corporation
-
- Baba Yoshinobu
- Department of Medicinal Chemistry, Faculty of Pharmaceutical Sciences, The University of Tokushima CREST, Japan Science and Technology Corporation
この論文をさがす
抄録
A laser-induced fluorescence detection system coupled with a highly sensitive silicon intensified target (SIT) video camera has been successfully applied to the imaging of a band for DNA fragment labeled with fluorescence dye and the visualizing of a separation process in the microchannel fabricated on a chip. We demonstrated that the imaging of the sample injection process is a powerful method for optimizing the electric potentials. Moreover, in microfabricated capillary electrophoresis (μ-CE) devices, continuous observation of the sample separation process is available. This direct imaging method is useful for the study of separation dynamics on μ-CE devices.
収録刊行物
-
- bioimages
-
bioimages 7 (4), 157-161, 1999
日本バイオイメージング学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1391975276374339328
-
- NII論文ID
- 10004718514
-
- NII書誌ID
- AA11084187
-
- ISSN
- 09192719
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可