Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer.

この論文をさがす

抄録

Gas chromatograph/mass spectrometer (GC/MS) and gas chromatograph/atomic emission detector (GC/AED) with temperature programmable pyrolyzer were used for analysis of pressure sensitive adhesives in adhesive tapes. Evolved gas analysis (EGA) profile of the adhesives was obtained by this temperature-programmed pyrolyzer directly coupled with MS via a deactivated metal capillary tube. The EGA profile suggested the optimal thermal desorption condition for additives and the subsequent optimal pyrolysis temperature for the remaining polymeric material. Rubbers were identified from pyrograms with the assistance of a new polymer library. The additives were selectively detected by monitoring hetero elements and these screening results of GC/AED enabled one to easily find additive peaks in complex chromatograms of GC/MS with interference of tackifier. The additives were able to be identified by the corresponding GC/MS data. Furthermore, the EGA profile gave significant data for crosslinking agents by using specific ion monitoring.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 16 (6), 627-631, 2000

    社団法人 日本分析化学会

被引用文献 (2)*注記

もっと見る

参考文献 (16)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ