内筒を通してCAG,PTCAが旋行できるIABP

書誌事項

タイトル別名
  • A new intraaortic balloon catheter with which coronary angiogaphy or percutaneous transluminal coronary angioplasty can be applied.

この論文をさがす

抄録

A new intraaortic balloon catheter is described in this paper. The balloon is a standard single chamber type and its volume is 35cc. The length of balloon catheter is 71cm. The outer annular lumen of the balloon catheter is 12F in size. The inner lumen of it is 6.5F in diameter and permits the insertion of a guiding catheter for coronary angiography (CAG) or percutaneous transluminal coronary angioplasty (PTCA). The bottom has a Y-shaped connector and is attached with a one-way valve Just as a regular sheath introducer. The material of the balloon is polyurethane. With only one puncture of femoral artery, the balloon is used for circulatory support during CAG or PTCA. We experienced five patients who were applied PTCA and one patient CAG, and no complications related to CAG, PTCA and IABP were observed.