- Clinical evaluation of infant membrane oxygenators, "Capiox 10R" in comparison with "MENOX AL-4000" and "D705 MIDIFLO".
この論文をさがす
抄録
We studied the clinical applicability of the newly developed membrane oxygenator Capiox SX-10R in comparison with the MENOX AL-4000 and the D705 MIDIFLO in open heart surgery on children. The Capiox SX-10R is a hollow fiber oxygenator with the detachable hardshell reservoir. Every oxygenator showed satisfactory O2 and CO2 transfer performance. There were no statistical differences in hemocompatibility among the three oxygenators in terms of the plasma free hemoglobin and platelet preservation ratios. As the priming volume of the Capiox SX-10R is 135ml, it seems to promise to reduce the volume wasted in blood transfusions and to enable pediatric open heart surgery without blood transfusion