新規購入インフュージョンポンプに対する導入時点検の重要性 Brand new infusion pumps may act faultily

この論文にアクセスする

この論文をさがす

著者

    • 大西 芳明 OHNISHI Yoshiaki
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine
    • 黒田 泰弘 KURODA Yasuhiro
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine
    • 福田 靖 FUKUTA Yasuhi
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine
    • 飯富 貴之 IITOMI Takayuki
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine
    • 佐藤 由美子 SATO Yumiko
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine
    • 阿部 正 ABE Tadashi
    • 徳島大学医学部附属病院救急部・集中治療部 Division of Intensive and Critical Care Medicine, Tokushima University School of Medicine

抄録

新規購入された輸液ポンプ(TE-172,TERUMO,東京)100台およびシリンジポンプ(TE-312C,TERUMO,東京)120台に対して,初期トラブルを未然に防止するために臨床工学技士による導入時点検を実施しメーカー仕様と比較した。輸液ポンプでは,その外観,セルフチェック動作,気泡検出機能には異常は認められなかった。閉塞検出機能では,1台で高圧域で閉塞警報が検出範囲内で作動せず調整を必要とした。流量精度試験では10台に異常が認められ,1台で低流量域で流量が著しく減少し調整を必要とした。バッテリー性能では,shut down時間には異常なかったが,1台で電圧低下警報の発生時間の短縮が認められバッテリーを交換した。シリンジポンプでは,その外観,セルフチェック動作,シリンジ検出機能,残量検出機能,閉塞検出機能,バッテリー電圧低下警報の発生時間には異常は認められなかった。流量精度試験では8台に異常値が得られたが調整を必要とするほどではなかった。点検には107.5時間が必要であった。臨床工学技士による機器購入時の導入時点検は,点検に時間を要するが,その性能を維持するために不可欠である。

We checked through one hundred brand-new infusion pumps (TE-172, Terumo, Tokyo) and one hundred and twenty brand-new syringe pumps (TE-312C, Terumo, Tokyo) before use to avoid unexpected incidents and compared our results with the company's specifications. No abnormality was found in infusion pumps in appearance, self-check function, and the air-in-the-line detecting function. One infusion pump showed its occulusion detecting function out of the limit at high pressure level (0.6-1.4kgf·cm<sup>-1</sup>) and was fixed. Ten infusion pumps showed inaccurate flow volumes at low flow level (25ml·hr<sup>-1</sup> for 60 minutes). We adjusted only one of them which had delivered extremely small quantity for its setting. Lifespan of a battery in any one of the infusion pumps was acceptable (231±47min) but a battery alarm of the infusion pump warned earlier (90min) than the standard. As for syringe pumps, there was no abnormality as well in appearance, self check function, and the functions to detect the size of syringes, the emptiness and the occulusion of the line. No abnormality was found in battery alarms and accuracy of flow rates. Although we found more than 3% aberration of flow rates in 8 pumps, it was in acceptable range and no adjustment was necessary. Working time to check all these pumps was 107.5 hours. We recommend that clinical engineers should appraise the functions of new medical equipments before operate them to the patients, even though it takes a great amount of time.

収録刊行物

  • 日本集中治療医学会雑誌 = Journal of the Japanese Society of Intensive Care Medicine

    日本集中治療医学会雑誌 = Journal of the Japanese Society of Intensive Care Medicine 9(2), 139-142, 2002-04-01

    The Japanese Society of Intensive Care Medicine