一様渦電流を用いた伝熱管渦流探傷用マルチバンクマルチコイルプローブ

書誌事項

タイトル別名
  • Multi-Bank Multi-Coil Probe for Heat Exchanger Tube Eddy Current Test using Uniform Eddy Current
  • イチヨウ ウズデンリュウ オ モチイタ デンネツカン カリュウタンショウヨウ マルチバンクマルチコイルプローブ

この論文をさがす

抄録

To enhance a wobbling noise reduction feature and 2D flaw resolution characteristic for tube eddy current ISI, the authors developed a new multi-bank multi-coil probe using uniform eddy current. The new probe has a large/long solenoid coil for excitation and small pancake pickup coils. The excitation coil induces a uniform eddy current that flows in a circumferential direction in the tube wall and the pickup coils are positioned in the area of the uniform eddy current. The self-differential feature makes the probe self-nulling and free of wobbing noise. The authors also developed a multi-bank probe that is constructed from four banks of coil. Each bank has eight pickup coils. All signals generated by the 32 pickup coils are integrated into a 2D defect signal just like a signal by an ordinary multi-coil probe. The multi-bank method allows easy establishment of a higher resolution multi-coil. The newly developed probe makes the tube test reliable with less noise and more information than a conventional multi-coil probe.

収録刊行物

  • 非破壊検査

    非破壊検査 53 (6), 366-372, 2004