3次元光コヒーレンストモグラフィ : OCTオフサルモスコープ 3 Dimensional Optical Coherence Tomography : OCT-Ophthalmoscope

この論文にアクセスする

この論文をさがす

著者

    • 上野 登輝夫 UENO Tokio
    • (株)ニデック 研究開発本部 技術開発部 光学開発課 NIDEK CO., LTD. Optic Development, Engineering Development Dept., R&D Div.

抄録

OCT-Ophthalmoscope scans the human retina in 2 dimensions by a galvo-scanner at high speed and constructs 2 dimensional OCT images (C-Scan images) by detecting the interference signal produced by the measuring light and the reference light. Furthermore, by moving the reference mirror along the optical axis, this system can construct 3 dimensional OCT images. The resolution of the system is 9 microns in the longitudinal direction and 18 microns in the transversal direction.<BR>It enables us by its unique C-Scan OCT images to observe retinal images not only in a one-dimensional intersection but also in the area or the volume and to diagnose or evaluate retinal disease objectively and numerically.