書誌事項
- タイトル別名
-
- Development of New Method for Discriminating Embolus from Air Bubble in Pulsatile Blood Flow Using Ultrasound
- チョウオンパ オ モチイタ ハクドウ ケツリュウチュウ ノ センシ ト キホウ ノ シキベツ ホウホウ ノ カイハツ
この論文をさがす
抄録
Patients who have operation for artificial valve substitution frequently suffer from brain embolism. Micro air bubbles are generated with opening and closing of an artificial valve in such a patient's blood. It is difficult to discriminate air bubbles from emboli by methods using ultrasonic wave. Hence, we tried to develop a new method for discriminating embolus from air bubble in pulsatile blood flow using ultrasound. We utilized the difference in motion of embolus and an air bubble when they are irradiated by ultrasound beam. It is considered that the difference in motion is caused from the difference in acoustic impedance and mass of them. As a result, we succeeded to discriminate emboli from micro air bubbles in pulsatile flows as well as in steady flow in the model experiments in vitro.
収録刊行物
-
- 可視化情報学会誌
-
可視化情報学会誌 26 (Supplement2), 97-98, 2006
社団法人 可視化情報学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679597472896
-
- NII論文ID
- 10018247262
- 40015522949
-
- NII書誌ID
- AA11453895
-
- ISSN
- 1884037X
- 09164731
-
- NDL書誌ID
- 8857691
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可