Evaluation of RF Heating on Humerus Implant in Phantoms during 1.5T MR Imaging and Comparisons with Electromagnetic Simulation

この論文にアクセスする

この論文をさがす

著者

    • MURANAKA Hiroyuki
    • Major in Medical Engineering and Technology, Graduate Course in Integrated Human Sciences Studies, Hiroshima International University
    • HORIGUCHI Takayoshi
    • Major in Medical Engineering and Technology, Graduate Course in Integrated Human Sciences Studies, Hiroshima International University
    • USUI Shuji
    • Major in Medical Engineering and Technology, Graduate Course in Integrated Human Sciences Studies, Hiroshima International University
    • UEDA Yoshitake
    • Major in Medical Engineering and Technology, Graduate Course in Integrated Human Sciences Studies, Hiroshima International University

抄録

Purpose: To evaluate the effect of radiofrequency (RF) heating on a metallic implant during magnetic resonance imaging (MRI), temperatures at several positions of an implant were measured, and results are compared with electromagnetic simulations using a finite element method.<br> Methods: A humerus nail implant made of stainless steel was embedded at various depths of tissue-equivalent gel-phantoms with loop (loop phantom) and partially cut loop (loop-cut phantom), and the phantoms were placed parallel to the static magnetic field of a 1.5T MRI device. Scans were conducted at maximum RF for 15 min, and temperatures were recorded with 2 RF-transparent fiberoptic sensors. Finally, electromagnetic-field analysis was performed.<br> Results: Temperatures increased at both ends of the implants at various depths, and temperature increase was suppressed with increasing depth. The maximum temperature rise was 12.3°C at the tip of the implant and decreased for the loop-cut phantom. These tendencies resembled the results of electromagnetic simulations.<br> Conclusion: RF heating was verified even in a nonmagnetizing metal implant in a case of excessive RF irradiation. Particularly, rapid temperature rise was observed at both ends of the implant having large curvatures. The difference in temperature increase by depth was found to reflect the skin-depth effect of RF intensity. Electromagnetic simulation was extremely useful for visualizing the eddy currents within the loop and loop-cut phantoms and for evaluating RF heating of a metallic implant for MRI safety.<br>

収録刊行物

  • Magnetic resonance in medical sciences : MRMS  

    Magnetic resonance in medical sciences : MRMS 5(2), 79-88, 2006-07-10 

    Japanese Society for Magnetic Resonance in Medicine

参考文献:  23件

参考文献を見るにはログインが必要です。ユーザIDをお持ちでない方は新規登録してください。

被引用文献:  6件

被引用文献を見るにはログインが必要です。ユーザIDをお持ちでない方は新規登録してください。

各種コード

  • NII論文ID(NAID)
    10018258529
  • NII書誌ID(NCID)
    AA11648770
  • 本文言語コード
    ENG
  • 資料種別
    ART
  • ISSN
    13473182
  • データ提供元
    CJP書誌  CJP引用  J-STAGE 
ページトップへ