Quantitative Analysis of Methacycline Hydrochloride by Direct Potentiometry Using the Internal Solid Contact Sensor

  • SUN Xian Xiang
    Department of Chemical Engineering, Jiangsu Polytechnic University
  • ABOUL-ENEIN Hassan Y.
    Pharmaceutical and Medicinal Chemistry Department, Pharmaceutical and Drug Industries Research Division, National Research Center (NCR)

Search this article

Abstract

An internal solid contact sensor (ISCS) for the determination of methacycline hydrochloride (MC·Cl), Pt/PPy/PVC(MC-PT), is described, based on the use of conducting poly(pyrrole) (PPy) as solid contact material and MC-phosphotungstate (PT) as the ion exchanger and dibutyl phthalate (DBP) as the plasticizer. A direct potentiometric method for the quantitative analysis of MC·Cl is also described. Under the condition of pH 2.7, the linear concentration range, slope (25°C) and detection limit of the sensor are 6.4 × 1.0-6 - 3.0 × 1.0-3 M, 52.4 ± 0.2 mV/decade and 4.4 × 1.0-6 M, respectively. The response time is <5 s. The determinations of MC·Cl in tablets were carried out by direct potentiometry. The average recovery and relative standard deviation are 100.1 and 0.7% (n = 4), respectively.

Journal

  • Analytical Sciences

    Analytical Sciences 23 (2), 231-233, 2007

    The Japan Society for Analytical Chemistry

References(14)*help

See more

Details 詳細情報について

Report a problem

Back to top