なぜリフレクトロンモードやマルチターンによって質量分解能が向上されるのか Why Can Reflectron and Multi-Turn Time-of-Flight Mass Spectrometer Improve Their Mass Resolving Powers?

この論文にアクセスする

この論文をさがす

著者

    • 田中 耕一 TANAKA Koichi
    • 株式会社島津製作所田中耕一記念質量分析研究所 Koichi Tanaka Mass Spectrometry Research Laboratory, Shimadzu Corporation
    • 吉野 健一 YOSHINO Ken-ichi
    • 神戸大学自然科学系先端融合研究環・バイオシグナル研究センター Biosignal Research Center, Organization of Advanced Science and Technology, Kobe University

抄録

Linear mode, reflectron mode, and multi-turn are techniques for <i>m</i>/<i>z</i> analysis in time-of-flight mass spectrometry. Reflectron mode and multi-turn enable analysis to be performed with a higher mass resolving power than that attained with linear mode. In this commentary, starting from the basic principles of time-of-flight mass spectrometry in linear mode, the possible improvements in mass resolving power that can be made by reflectron mode and multi-turn techniques will be explained.