Application of H-point standard addition method and partial least squares to the simultaneous kinetic-potentiometric determination of hydrazine an phenylhydrazine

書誌事項

タイトル別名
  • Application of H-Point Standard Addition Method and Partial Least Squares to the Simultaneous Kinetic-Potentiometric Determination of Hydrazine and Phenylhydrazine

この論文をさがす

抄録

Simultaneous determination of hydrazine (HZ) and phenylhydrazine (PHZ) by H-point standard addition method (HPSAM) and partial least squares (PLS) regression was carried out based on kinetic data from novel potentiometry methods. The rate of chloride ion production in the reaction of HZ and PHZ with N-chlorosuccinimide (NCS) was monitored by a chloride ion-selective electrode. The experimental data show the good ability of ion-selective electrodes (ISEs) as detectors not only for the direct determination of chloride ion but also for simultaneous kinetic-potentiometric analysis using HPSAM and PLS methods. The methods are based on the differences observed in the production rate of chloride ions. The results show that simultaneous determination of HZ and PHZ can be performed in concentration ranges of 0.5 - 20.0 and 0.8 - 25.0 µg mL-1, respectively. The total relative standard error for applying the PLS method to 8 synthetic samples in the concentration ranges of 1.0 - 16.0 µg mL-1 for HZ and 2.0 - 16.0 µg mL-1 for PHZ was 3.96. In order for the selectivity of the method to be assessed, we evaluated the effects of certain foreign ions upon the reaction rate and assessed the selectivity of the method. Both methods (PLS and HPSAM) were evaluated using a set of synthetic sample mixtures and then applied for simultaneous determination of HZ and PHZ in water samples.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 24 (2), 261-266, 2008

    社団法人 日本分析化学会

参考文献 (80)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ