Abstract

Medical devices have been obliged to satisfy electromagnetic compatibility by revision of the pharmaceutical affairs law. However, even if the medical devices satisfy the electromagnetic compatibility based on the law, it is not necessarily safe. Sometimes, malfunctions of cardiac pacemaker are caused by the magnetic field leaked from an induction heating cooker. In this paper, a new method of electromagnetic susceptability (EMS) evaluation is proposed, and a loop coil for the magnetic field immunity test in the frequency range from 10kHz to 3MHz is designed and developed. As a result, the loop coil made on an experimental basis generated uniform magnetic field with a fluctuation within 3.3dB in the loop coil pane and 5.6dB along the coil axis.