Trace Analysis of Released Metallic Ions in Static Immersion Test for Characterization of Metallic Biomaterials

Bibliographic Information

Other Title
  • Trace Analysis of Released Metallic Ions in Static Immersion Test for Characterlization of Metallic Biomaterials

Search this article

Abstract

For characterization of corrosion resistance of metallic biomaterials, determination of trace amounts of metallic ions released from the materials in using static immersion into simulated body fluids (SBF) were investigated. In using a H2SO4-fume pre-treatment method, sensitive, precise and accurate determination of the trace metallic elements in SBFs could be performed by ICP-OES. For accurate analysis, it was necessary to employ a matrix-matched solution for the calibration. Moreover, usage of a PFA vessel in the static immersion test was recommended for the prevention of contamination. Thus, it was possible to determine μg dm−3 (ppb) order of elements in SBFs and evaluate nano-gram order of the released metallic ions.

Journal

  • MATERIALS TRANSACTIONS

    MATERIALS TRANSACTIONS 49 (6), 1342-1345, 2008

    The Japan Institute of Metals and Materials

Citations (2)*help

See more

References(23)*help

See more

Details 詳細情報について

Report a problem