二段式流量分配法による分圧真空計の校正サービス Calibration Service for Partial Pressure Analyzers by Two-stage Flow-dividing Method

この論文にアクセスする

この論文をさがす

著者

抄録

  The calibration service for partial pressure analyzers (PPA) has been carried out from April in 2007 at NMIJ/AIST using a two-stage flow-dividing system. The pressure range is from 10<sup>-6</sup> to 10<sup>-4</sup> Pa and the test gases are N<sub>2</sub>, Ar, He and H<sub>2</sub>.<br>   Results of the calibration for two types of PPA were also shown. The sensitivities of both PPAs were almost constant in the pressure range from 10<sup>-6</sup> to 10<sup>-4</sup> Pa, while they changed significantly with pressure higher than 10<sup>-4</sup> Pa. The calibration uncertainties mainly caused by the instability of PPA indication were about 7% (confidence level 95%). The stability of sensitivities of both PPAs was in about 15% for 7 months.<br>