Rapid enumeration of bacteria in the production process of RO water in a dialysis facility by fluorescent staining methods
-
- Naramura Tomotaka
- Ide Clinic
-
- Sato Kazuhiro
- Ide Clinic
-
- Matsuura Hiromi
- Ide Clinic
-
- Tokura Hideomi
- Ide Clinic
-
- Inoguchi Aki
- Ide Clinic
-
- Koga Hisae
- Ide Clinic
-
- Ide Takao
- Ide Clinic
Bibliographic Information
- Other Title
-
- 蛍光染色法を用いたRO水製造工程中に存在する細菌の迅速評価
- ケイコウ センショクホウ オ モチイタ ROスイ セイゾウ コウテイチュウ ニ ソンザイスル サイキン ノ ジンソク ヒョウカ
Search this article
Abstract
Bacterial numbers during each of 5 steps in the production process of RO water, which dialysis fluid was made from, were measured using fluorescent staining methods (DAPI and 6CFDA staining, and microcolony methods). And measured values were compared with those obtained by conventional methods (existing culture method and measurement of endotoxin activity). According to the results of the comparison, we confirmed that bacterial number in the production process of RO water was not measured precisely by the existing culture methods and combination with fluorescent staining methods was effective for monitoring, as indicated in related reports. As a result of the measurements of the bacterial number at these 5 stages using various methods, we confirmed that the bacterial number required attention at these points from tap water to RO tank in order to ensure the safety of dialysis fluid.
Journal
-
- Nihon Toseki Igakkai Zasshi
-
Nihon Toseki Igakkai Zasshi 40 (12), 1051-1056, 2007
The Japanese Society for Dialysis Therapy
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679654415104
-
- NII Article ID
- 10021274203
-
- NII Book ID
- AN10432053
-
- ISSN
- 1883082X
- 13403451
-
- NDL BIB ID
- 9318408
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed