PILOT PRODUCTION OF THE WEDGE FILTER FOR THE TBI (TOTAL BODY IRRADIATION)

  • IKEZAKI Hiromi
    Department of Radiological Technology, Kyorin University Hospital
  • IKEDA Ikuo
    Department of Radiological Technology, Kyorin University Hospital
  • MARUYAMA Yasushi
    Department of Radiological Technology, Kyorin University Hospital
  • NAKO Yasunobu
    Department of Radiological Technology, Kyorin University Hospital
  • TONARI Ayako
    Department of Radiology, Kyorin University School of Medicine
  • KUSUDA Junko
    Department of Radiology, Kyorin University School of Medicine
  • TAKAYAMA Makoto
    Department of Radiology, Kyorin University School of Medicine

Bibliographic Information

Other Title
  • 全身照射用ウェッジフィルタの作成

Search this article

Abstract

Total body irradiation (TBI) is performed by various methods, such as a long SSD method and a translational couch method. For patient safety in carrying out TBI, the patient should be placed on the supine position and prone position near the floor. TBI is performed from 2 opposite ports (AP/PA) with a lenear accelerator (10 MV X-ray). We experiemented with a wedge filter for TBI created by us,, which makes dose distribution to a floor uniform. The wedge filter, made of iron alloy, was attached to the linear accelerator. In designing the wedge filter, thickness of the lead-made wedge filter can be calculated numerically from the ratio of linear attenuation coefficient of iron alloy and lead. In measuring the dose profile for a phantom of 20 cm thick, dose homogeneity less than 10% was proved by the wedge filter for TBI. (Article in Japanese)

Journal

  • The Journal of JASTRO

    The Journal of JASTRO 19 (4), 253-258, 2007

    Japanese Society for Therapeutic Radiology and Oncology

References(8)*help

See more

Details 詳細情報について

Report a problem