超伝導重力計連続観測に対する液体ヘリウムレベル変化の影響?南極・昭和基地での観測例?

書誌事項

タイトル別名
  • Influence of Liquid Helium Level Changes on Continuous Gravity Observation with Superconducting Gravimeter at Syowa Station, Antarctica

この論文をさがす

抄録

Change of the liquid helium level affects the position change of the proof mass of superconducting gravimeter (SG). We estimated its effect on three different types of SGs (TT-70#016, CT#043, and OSG#058) by approximating the spatial relationship between the SG sensor unit and liquid helium reservoir using a double-layered cylinder model where the inner cylinder and outer one are assumed SG sensor unit and liquid helium reservoir tank, respectively.<BR> From these calculations, it was predicted for SG TT-70#016 that a gravity change of 0.88 µGal (1 µGal=10-8m•s-2) results when the liquid helium level in the reservoir is increased from 0% to 100%. Similar gravity changes of 0.25 µGal and 0.35 µGal were also predicted for SGs CT#043 and OSG#058, respectively. In addition, artificially-induced, rapid gravity changes of approximately 0.9 µGal over the course of two hours likely occur during the filling of SGs with liquid helium from 20% to 100% for TT-70#016. Actual gravity changes correlated with the rapid liquid helium level changes were also found in the observed gravity record of SG CT#043 in 2003, although the actual gravity changes were greater than the predicted ones.<BR> Gravity changes of approximately 0.1 µGal due to fluctuations of 5% to 14% in the liquid helium level change are also possible during the normal continuous observation with SGs. It is therefore preferable to monitor the liquid helium level routinely when measuring small gravity signals in the order of sub-micro Gal to correct for these effects.

収録刊行物

被引用文献 (1)*注記

もっと見る

参考文献 (6)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ