回転パノラマ撮影装置を用いた顎関節撮影における吸収線量と実効線量

書誌事項

タイトル別名
  • The Absorbed Dose and the Effective Dose of Panoramic Temporo Mandibular Joint Radiography
  • カイテン パノラマ サツエイ ソウチ オ モチイタ ガク カンセツ サツエイ ニ オケル キュウシュウセンリョウ ト ジッコウセンリョウ

この論文をさがす

抄録

Purpose: This study measured the radiation doses absorbed by the patient during Panoramic temporo mandibular joint radiography (Panoramic TMJ), Schüllers method and Orbitoramus projection. The dose of the frontal view in Panoramic TMJ was compared to that with Orbitoramus projection and the lateral view in Panoramic TMJ was compared to that with Schüllers method. Methods: We measured the doses received by various organs and calculated the effective doses using the guidelines of the International Commission on Radiological Protection in Publication 103. Organ absorbed doses were measured using an anthropomorphic phantom, loaded with thermoluminescent dosimeters (TLD), located at 160 sensitive sites. The dose shows the sum value of irradiation on both the right and left sides. In addition, we set a few different exposure field sizes. Result: The effective dose for a frontal view in Panoramic TMJ was 11 μSv, and that for the lateral view was 14 μSv. Conclusion: The lens of the Orbitoramus projection was 40 times higher than the frontal view in Panoramic TMJ. Although the effective dose of the lateral view in Panoramic TMJ was 3 times higher than that of the small exposure field (10×10 cm on film) in Schüller’s method, it was the same as that of a mid-sized exposure field. When the exposure field in the inferior 1/3 was reduced during panoramic TMJ, the effective doses could be decreased. Therefore we recommend that the size of the exposure field in Panoramic TMJ be decreased.

収録刊行物

参考文献 (15)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ