積分カフ脈波拡張率に基づく血管内皮機能の非観血評価

書誌事項

タイトル別名
  • Noninvasive Evaluation of Endothelial Function Based on Dilation Rate of Integrated Air-Cuff Plethysmogram
  • セキブン カフ ミャクハ カクチョウリツ ニ モトズク ケッカン ナイヒ キノウ ノ ヒカン ケツ ヒョウカ

この論文をさがす

抄録

This paper proposes a method for evaluating endothelial function based on the dilation rate of an integrated air-cuff plethysmogram measured using an oscillometric approach. The method is not affected by blood pressure variations, and can be used to assess endothelial function quickly. In the study, the dilation rate of an integrated air-cuff plethysmogram was simulated to verify that the proposed method is suitable for evaluating arterial compliance changes related to endothelial function. The simulation results showed that changes in arterial viscoelastic characteristics can be assessed using the technique. Then, the proposed dilation rate for the integrated air-cuff plethysmogram and the value of percent flow-mediated vasodilation (%FMD) (a standard index used to evaluate endothelial function) were measured in the same subjects, with the results showing a significant correlation (r= 0.589) between the two. In addition, receiver operation characteristics (ROC) analysis was conducted to verify the method's effectiveness in determining the risk of arteriosclerosis. The results showed that the area under the curve (AUC) of the proposed index was 0.914, which was greater than or equal to the %FMD value. It was therefore concluded that the proposed method has potential for clinical use.

収録刊行物

  • 医療機器学

    医療機器学 82 (3), 259-266, 2012

    一般社団法人日本医療機器学会