- 小林 裕之
- 大阪工業大学工学部
書誌事項
- タイトル別名
-
- EMG/ECG Acquisition System with Online Adjustable Parameters Using ZigBee Wireless Technology
- ドウテキ ニ パラメタ オ ヘンコウ カノウ ナ ZigBee オ モチイタ ムセン EMG/ECG シュトク システム
この論文をさがす
抄録
This paper deals with a novel wireless bio-signal acquisition system employing ZigBee wireless technology, which consists of mainly two components, that is, intelligent electrode and data acquisition host. The former is the main topic of this paper. It is put on a subject's body to amplify bio-signal such as EMG or ECG and stream its data at upto 2 ksps. One of the most remarkable feature of the intelligent electrode is that it can change its own parameters including both digital and analog ones on-line. The author describes its design first, then introduces a small, light and low cost implementation of the intelligent electrode named as “VAMPIRE-BAT.” And he show some experimental results to confirm its usability and to estimate its practical performances.
収録刊行物
-
- 電気学会論文誌C(電子・情報・システム部門誌)
-
電気学会論文誌C(電子・情報・システム部門誌) 132 (5), 632-639, 2012
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390001204609518976
-
- NII論文ID
- 10030605887
-
- NII書誌ID
- AN10065950
-
- BIBCODE
- 2012ITEIS.132..632K
-
- ISSN
- 13488155
- 03854221
-
- NDL書誌ID
- 023777103
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可