書誌事項
- タイトル別名
-
- ヘイコウバン ネンドケイ ノ シサク
- <Original Papers>Parallel Plate Viscometer and Its Characteristic
この論文をさがす
抄録
application/pdf
A parallel plate viscometer was constructed by utilizing the combination of linear variable transformer and a microcomputer. The characteristic of this viscometer was determined by measurement of viscosity of a NBS standard glass 711. The experimental results show that the applicability of this viscometer for the viscosity range from 10^6 to 10^<10> poises. There are two methods of measurement using the parallel plate viscometer. Namely, one is the static method in which temperature stays constant during each measurement and the other is the dynamic method in which temperature varies continuously through the measurement. From the comparison of two methods, it is concluded that the dynamic method gives a higher value of viscosity than that of the static method. The experimental errors due to temperature, the expansion of the mullite reference rod, and the expansion of the sample are discussed.
紀要類(bulletin)
619514 bytes
収録刊行物
-
- 東北大學選鑛製錬研究所彙報 = Bulletin of the Research Institute of Mineral Dressing and Metallurgy, Tohoku University
-
東北大學選鑛製錬研究所彙報 = Bulletin of the Research Institute of Mineral Dressing and Metallurgy, Tohoku University 38 (1), 1-10, 1982-10-20
東北大學選鉱製錬研究所
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1050001202743432576
-
- NII論文ID
- 110000202801
-
- NII書誌ID
- AN00168425
-
- ISSN
- 0040876X
-
- HANDLE
- 10097/32824
-
- NDL書誌ID
- 2454230
-
- 本文言語コード
- ja
-
- 資料種別
- departmental bulletin paper
-
- データソース種別
-
- IRDB
- NDL
- CiNii Articles