ペルオキシダーゼ固定電極及びポリエチレングリコール修飾酵素を用いる血清中高密度リポタンパクコレステロール及び総コレステロールの直接測定

書誌事項

タイトル別名
  • Direct determination of high-density lipoprotein- and total cholesterol in serum using a peroxidase-entrapped electrode and polyethylene glycol-modified enzymes.
  • ペルオキシダーゼ コテイ デンキョク オヨビ ポリエチレン グリコール シュウ

この論文をさがす

抄録

The total concentrations of lipoprotein (LP) cholesterol (CR) in serum were directly determined by amperometric measurements of H2O2 generated by polyethylene glycol (PEG)-modified cholesterol esterase and cholesterol oxidase at a membrane-covered, peroxidase-entrapped and ferrocene-embedded carbon-paste electrode. The concentrations of LP aggregators, such as sodium dextran sulfate and MgCl2, and the amount of the PEG-modified enzymes were optimized to determine the high-density lipoprotein (HDL)-CR concentration. The characteristics of the discrimination between HDL and other LPs by the PEG-modified enzymes and the aggregators are discussed. Using a surfactant as a solubilizer of LP aggregation, the HDL-and total CR concentrations have been discriminatively determined. This method was applied to the determination of the HDL-CR values of 17 human samples. The evaluated values were well correlated to those determined by a commercial spectrophotometric determination kit using the PEG-modified enzymes and also by an amperometric determination of the HDL-fraction prepared by an ordinary precipitation method.

収録刊行物

  • 分析化学

    分析化学 47 (4), 233-238, 1998

    公益社団法人 日本分析化学会