6,9-ジクロロ-2-メトキシアクリジンを蛍光試薬とする塩化水素ガス定量用テープの開発

書誌事項

タイトル別名
  • Development of a monitoring tape for fluorescence detection of hydrogen chloride gas using 6,9-dichloro-2-methoxyacridine.
  • 6,9‐ジクロロ‐2‐メトキシアクリジンを蛍光試薬とする塩化水素ガス定量用テープの開発
  • 6 9 ジクロロ 2 メトキシアクリジン オ ケイコウ シヤク ト スル エン

この論文をさがす

抄録

A porous cellulose tape coated with silica gel as an absorbent and impregnated with a processing solution which includes a fluorescent pH indicator (6, 9-dichloro-2-methoxyacridine), p-toluenesulfonic acid, glycerin and methanol is a highly sensitive means of detecting hydrogen chloride gas in air. When a sample containing hydrogen chloride was passed through the tape, the pH of the tape decreased. The fluorescence intensity was proportional to the concentration of hydrogen chloride gas at a constant sampling time and flow rate. The fluorescence intensities were measured at 375nm(λex)and 510nm(λem). The detection limit was 0.05ppm for hydrogen chloride gas at a sampling time of 60 s and a flow rate of 400ml/min. No interference was observed from ethanol (1v/v%), trichloroethylene (1v/v%), acetone (1v/v%), carbon dioxide (4.9viv%), nitrogen dioxide (100ppm), sulfur dioxide (50ppm), acetic acid gas (24ppm), or hydrogen fluoride (10ppm).