抄録

Absorbed dose was measured in clinical X-ray examinations using TLD. Moreover, we distributed the levels of radiation exposure into 3 classes. The presumed dose of the internal organs, e. g., uterus dose, was computed to depth doses with a surface dose. This information provides a prediction of the influence of radiation, and the examination can be performed with te informed consent of the patient. Moreover, we examined the distribution of the level of absorbed dose. We proposed two kinds of radiation exposure level, one to the fetus in a pregnant woman and a general level of radiation exposure that is not applied to pregnant women. The levels were as follows : 0.5 mGy and 100 mGy were considered the boundaries for fetal radiation exposure in a pregnant woman, and 200 mGy and 3 Gy were considered the boundaries for the general level of radiation exposure (excluding pregnant women).