マルチスライスCTにおける画質の基礎物理評価

書誌事項

タイトル別名
  • Evaluation of Physical Characteristics in Multi-slice CT

この論文をさがす

抄録

This paper describes the results of a quality-of-image evaluation in multi-slice CT equipment. Image noise, section sensitivity profile (SSP), and spatial resolution of axial plane images were determined using two types of multi-slice CT equipment (Hispeed Advantage QX/I:GE-YMS Co. Ltd.; Aquilion: Toshiba Co. Ltd.). Suggestion: Based on theoretical considerations image noise was increased for helical pitch. Spatial resolution in the axial plane has been improved in the frequency domain by quarter-quarter offset reconstruction. Although slice thickness changed with helical pitch, as compared with a single CT, the profile has been improved notably. The limits of thin stratification of beam collimation and SSP evaluation by the bead method (point spread function using a small tungsten ball)were examined experimentally.