書誌事項
- タイトル別名
-
- Evaluation of Physical Characteristics in Multi-slice CT
この論文をさがす
抄録
This paper describes the results of a quality-of-image evaluation in multi-slice CT equipment. Image noise, section sensitivity profile (SSP), and spatial resolution of axial plane images were determined using two types of multi-slice CT equipment (Hispeed Advantage QX/I:GE-YMS Co. Ltd.; Aquilion: Toshiba Co. Ltd.). Suggestion: Based on theoretical considerations image noise was increased for helical pitch. Spatial resolution in the axial plane has been improved in the frequency domain by quarter-quarter offset reconstruction. Although slice thickness changed with helical pitch, as compared with a single CT, the profile has been improved notably. The limits of thin stratification of beam collimation and SSP evaluation by the bead method (point spread function using a small tungsten ball)were examined experimentally.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 56 (12), 1447-1453, 2000