- Studies on Stress Distribution of Bone-bonded and Bone-adapted Titanium Endosseous Implants and Surrounding Bone. Part 2. Effect of Countersink Angle Variation to Stress Distribution in Bone.
- Part 2. Effect of Countersink Angle Variation to Stress Distribution in Bone
- 第2報カウンターシンク角度が周囲骨内応力に及ぼす影響
この論文をさがす
抄録
The purpose of this study was to investigate stress distribution in bone around bone-bonded and bone-adapted titanium cylindrical implants with various countersink angles using 2-dimensional finite element analysis. In the bone-adapted model, the slip element was used at the interface between the bone and implant. The following results were obtained.<BR>1. Stresses were observed at the compact bone around the countersink in both bone-bonded and boneadapted implants. Nevertheless, stress concentration differed in the location. In bone-bonded implants, the stress concentration was observed at the upper most part of the compact bone around the countersink of the implants. Whereas, in bone-adapted implants it was observed at the compact bone around the edge of the countersink.<BR>2. In bone-bonded implants, various countersink angles scarcely affected the stress value in bone. But in bone-adapted implants, the 75 degrees of countersink angle reduced the stress value utmost.
収録刊行物
-
- 日本補綴歯科学会雑誌
-
日本補綴歯科学会雑誌 39 (5), 844-850, 1995