書誌事項
- タイトル別名
-
- Mechanical Properties of Mesh-Type Artificial Tracheas
- メッシュガタ ジンコウキカン ノ リキガク トクセイ
この論文をさがす
抄録
Many studies have been made to develop artificial tracheas with good biocompatibility and mechanical performance. We have designed a new type of mesh-type artificial tracheas. The basic frame of the specimens was composed of a mesh cylinder and stent attached to the outside of the cylinder. We have demonstrated from the mechanical tests that artificial tracheas with ring-type stents are stiffer than those with spiral-type stents. Artificial tracheas with arc-shaped shaped stents are half as stiff as artificial tracheas with whole ring stents. We have compared the mechanical properties of mesh-type artificial tracheas with nylon ring stents and those of canine tracheas, and have found the proper number of rings per unit length for the mesh-type artificial tracheas.
収録刊行物
-
- 材料
-
材料 54 (1), 85-89, 2005
公益社団法人 日本材料学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680394598400
-
- NII論文ID
- 110006266388
-
- NII書誌ID
- AN00096175
-
- ISSN
- 18807488
- 05145163
-
- NDL書誌ID
- 7221047
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可