- Bedside Chest Radiography and Optimization of Radiographic Conditions in the NICU with the Fuji Computed Radiography System
- FCR オ モチイタ NICU ビョウシツ サツエイ ニ オケル キョウブ ケイカ カンサツ ト サツエイ ジョウケン ノ サイテキカ
この論文をさがす
抄録
We devised an application that uses the “Maximum value reading method (AutoIV)” for bedside chest radiography in a neonatal intensive care unit (NICU) that used the Fuji Computed Radiography (FCR) System. The application, named AutoIV-N, uses the relationship for density correction between radiographic conditions (mAs) and the Display Parameter (GS). GS=f (mAs) can be considered the relationship that connects FCR and the X-ray generator. When AutoIV-N is used, radiographic image contrast does not change. Further, radiographic image density fluctuation can be eliminated by random elements such as X-ray output fluctuation of the X-ray generator and the decline of photo-stimulated luminescence caused by fading of the imaging plate. Accordingly, image recording that is suitable for follow-up chest radiography is made possible. We choose nine patients and performed a comparison of radiographic density fluctuation in AutoIV-N and Fix. AutoIV-N was found to be more stable than Fix. It is possible to use the radiographic imaging condition that is optimized for all patients in the NICU by AutoIV-N. This facilitates radiation exposure optimization in medicine.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 64 (1), 15-24, 2008
公益社団法人 日本放射線技術学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001206361492096
-
- NII論文ID
- 110006548883
-
- NII書誌ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 9353574
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可