人工心臓用経皮伝送システムにおける電力伝送範囲拡大に関する検討

書誌事項

タイトル別名
  • Examination to Improve Transmissible Range in the Transcutaneous Energy Transmission System for the Artificial Heart
  • ジンコウ シンゾウヨウ ケイヒ デンソウ システム ニ オケル デンリョク デンソウ ハンイ カクダイ ニ カンスル ケントウ

この論文をさがす

抄録

Transcutaneous Energy Transmission System (TETS) has an advantage to send power inside the body without penetrating to the skin by a wire. It does not only prevent the patients from infection-associated syndromes but also it aims at improvement of their's Quality Of Life (QOL). However, TETS is weak a magnetic coupling, and becomes a problem of the coil heating from the power loss. Moreover, TETS has a difficulty that it has to operate under an unstable coupling factor and varying load resistance. In this paper, we examined the method of obtaining a stable output voltage for TETS system according to kinds of the artificial heart.