G020013 人工関節置換膝における膝蓋大腿関節面の接触圧力評価([G02001]バイオエンジニアリング部門一般セッション(1))

書誌事項

タイトル別名
  • G020013 Estimation of Contact Pressure on Patellofemoral Joint for Artificial Knee Implants

抄録

After the total knee arthroplasty, it is important to evaluate between the patella component and the femoral component, because patella components affect the complications and the flexion angle. The purpose of this study was to investigate the mechanisms of contact between the patellofemoral joint. Femur and tibia were created by aluminum alloy. The implanted knee model was constructed to replace the cruciate-retaining artificial knee joint to them. Artificial ligaments were used as quadriceps model and patellar tendon model. The contact pressure were measured at seven flexion angle (90 degree, 100 degree, 110 degree, 120 degree, 130degree, 140degree and 150 degree) under the applied loads of 1000N. In result, the contact pressure was considered to increase with rising flexion angle, but such a phenomenon was not showed. It is possible that not increased contact pressure is effect of contact between femur and quadriceps. Understanding the advantage and disadvantage of the patella component function may help to develop sophisticated component design.

収録刊行物

  • 年次大会

    年次大会 2011 (0), _G020013-1-_G020013-3, 2011

    一般社団法人 日本機械学会