Hemolytic Anemia soon after Replacement of Ascending Aorta for acute Aortic Dissection

この論文をさがす

抄録

type:Case Report

A 61-year-old female underwent replacement of the ascending aorta for Type A aortic dissection, then gradually developed severe hemolytic anemia over 1 month. The cause of hemolysis was found to be mechanical damage of red blood cells caused by an inverted felt strip at the proximal anastomosis. A reoperation for resection of the felt and repair of the proximal anastomosis successfully resolved the problem. Here, we report a rare case of hemolytic anemia at the site of an inverted felt strip that occurred after replacement of the ascending aorta.