Read/Search this Article
Abstract
Urinary 8-hydroxy-2'-deoxyguanosine (8-OHdG) is a widely used noninvasive biomarker of oxidative stress. A selective, sensitive and rapid method for determining 8-OHdG in human urine was developed using hydrophilic interaction chromatography-tandem mass spectrometry (HILIC-MS/MS) with electrospray ionization. 8-OHdG and isotopically labeled 8-OHdG (internal standard) were separated on a HILIC column with a mobile phase of 10. mM ammonium acetate: acetonitrile (1:9, v/v) within 10. min and detected by using a positive electrospray ionization interface under the selected reaction monitoring mode. The detection limits of 8-OHdG (corresponding to a signal-to-noise ratio of 3) for the HILIC-MS/MS system and the conventional method using a reversed-phase column with MS/MS were 1.0 and 26.0. fmol/injection, respectively. The proposed method makes it possible to monitor the basal level of urinary 8-OHdG from non-exposed healthy subjects and can be used for large-scale human studies. © 2012 Elsevier B.V.
Journal
- Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
-
Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences 893-894, 173-176, 2012-04-15
Elsevier B.V.
Share