3入力指標によるファジィ・レート制御

DOI

書誌事項

タイトル別名
  • Fuzzy control of multisensor cardiac pacemaker with 3 parameters.

抄録

We have reported the fuzzy-rates control by 2 parameters (minute ventilation, MV and oxygen saturation, OS) in rate adaptive cardiac pacing. Although the overall agreement between fuzzy-rates and target rates was satisfactory, the fuzzy-rates before and after exercise were almost at the same level (fuzzy-rates immediately decreased to the restinglevel before exercise, as far as MV and OS are used as the rate-controlling parameters, it is impossible to keep the fuzzy-rates higher after exercise because MV and OS are almost the same level before and after exercise. To improve this, we propose to add one more parameter which has generally the different values before and after exercise. We developed the automatic generation method of control rules for the 3 parameters pacing system and demonstrated its feasibility using MV, OS, and temperature, as the 3rd parameter. The preliminary simulation results were satisfactory. Automatic tuning methods of membershi functions and controlrules are under study.

収録刊行物

  • 人工臓器

    人工臓器 23 (1), 199-202, 1994

    一般社団法人 日本人工臓器学会

詳細情報 詳細情報について

  • CRID
    1390001204957170432
  • NII論文ID
    130001732383
  • DOI
    10.11392/jsao1972.23.199
  • ISSN
    18836097
    03000818
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘