体外循環のPH,酸素分圧,炭酸ガス分圧の連続的モニター法に関する研究

DOI

書誌事項

タイトル別名
  • Continuous Po2, Pco2 and PH monitoring during extracorporeal circulation.

この論文をさがす

抄録

A blood gas monitoring system during extracorporeal circulation was made and studied. This system was composed of a Po2 electrode (mnt Po2), Pco2 electrode (mnt Pco2) and PH electrode (mnt PH). Two systems were to monitor blood gas during ECC, one for arterial blood and the other for venous blood. The value of correlation coefficient between mnt Po2 and Po2 was 0.987 for 220 points, that of mnt Pco2 was 0.903 for 156 points and that mnt PH and PH was 0.959 for 82 points. This ECC monitoring system is believed to be much promissing.

収録刊行物

  • 人工臓器

    人工臓器 16 (1), 494-497, 1987

    一般社団法人 日本人工臓器学会

詳細情報 詳細情報について