Design of an implanted compact antenna for an artificial cardiac pacemaker system
-
- Lee Soonyong
- Department of Electrical and Computer Engineering, Hanyang University
-
- Seo Wonbum
- Department of Electrical and Computer Engineering, Hanyang University
-
- Ito Koichi
- Graduate School of Engineering, Chiba University
-
- Choi Jaehoon
- Department of Electrical and Computer Engineering, Hanyang University
抄録
An implanted compact antenna for an artificial cardiac pacemaker is proposed. The dimension of the pacemaker system, including the antenna element, is 30mm × 35mm × 7mm. When the antenna is embedded in a semi-solid flat phantom with equivalent electrical properties as the human body, S11 value is -19.2dB at 403.5MHz and the -10dB impedance bandwidth of the antenna is 10MHz (399∼409MHz). The proposed antenna in the phantom has a peak gain of -24.61dBi at 403.5MHz. The measured specific absorption ratio (SAR) value of the proposed antenna is 0.0079W/Kg (1g tissue). Moreover, to estimate the communication performance of the proposed antenna operated in the real environment, a link budget analysis is performed.
収録刊行物
-
- IEICE Electronics Express
-
IEICE Electronics Express 8 (24), 2112-2117, 2011
一般社団法人 電子情報通信学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680188522368
-
- NII論文ID
- 130001920415
-
- ISSN
- 13492543
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可